Quantification of systematic errors in the electron density and temperature measured with Thomson scattering at W7-X
Abstract
The electron density and temperature profiles measured with Thomson scattering at the stellarator Wendelstein 7-X show features which seem to be unphysical, but so far could not be associated with any source of error considered in the data processing. A detailed Bayesian analysis reveals that errors in the spectral calibration cannot explain the features observed in the profiles. Rather, it seems that small fluctuations in the laser position are sufficient to affect the profile substantially. The impact of these fluctuations depends on the laser position itself, which, in turn, provides a method to find the optimum laser alignment in the future.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- September 2023
- DOI:
- 10.1088/1748-0221/18/09/P09019
- arXiv:
- arXiv:2111.03562
- Bibcode:
- 2023JInst..18P9019N
- Keywords:
-
- Analysis and statistical methods;
- Detector alignment and calibration methods (lasers;
- sources;
- particle-beams);
- Pattern recognition;
- cluster finding;
- calibration and fitting methods;
- Plasma diagnostics - interferometry;
- spectroscopy and imaging;
- Physics - Plasma Physics;
- Physics - Instrumentation and Detectors