External upgrades to improve the RV precision of the APOGEE Spectrographs
Abstract
Several external hardware upgrades have been developed for the APOGEE Spectrographs as part of the Sloan Digital Sky Survey-V (SDSS-V) to improve their radial velocity (RV) precision from a floor of 100-200 m/sec to approx. 30 m/sec. The upgrades include: (1) Back Pressure Regulator (BPR) systems to stabilize the internal instrument LN2 tank boil-off pressure, lessening induced movement of the APOGEE optical bench; (2) Fabry-Perot Interferometer (FPI) calibration sources to improve wavelength calibration; and (3), the use of octagonal core fiber segments in the fiber train to improve radial scrambling. We discuss the fabrication, commissioning, and early performance of these upgrades.