Dose-area product in pediatric barium meal procedures
Abstract
Pediatric barium meal procedures result in a higher patient dose than conventional X-ray imaging. Due to the fact that pediatric patients are more radiosensitive and more likely to develop mutations because of the radiation exposure during their lifetime a high patient dose is a concerning issue. The aim of this work was to test the applied earlier dose-area evaluation method comparing with direct DAP measurements and to see the changes in patient dose with the acquisition of a new, more modern, equipment. The direct measurements with a DAP meter confirmed the dose evaluation method used in our earlier works and the possibility of dose reduction shown there. Implementation of a modern equipment with new features enabling dose reduction does not guarantee a positive result without staff training and even standardization of the technical parameters with subsequent demand to follow them.
- Publication:
-
Radiation Physics and Chemistry
- Pub Date:
- February 2019
- DOI:
- 10.1016/j.radphyschem.2018.07.019
- Bibcode:
- 2019RaPC..155...53M
- Keywords:
-
- Pediatric fluoroscopy;
- Barium meal;
- DAP measurements