On HMI's Mod-L Sequence: Test and Evaluation
Abstract
HMI Mod-L sequence can produce full Stokes parameters at a cadence of 90 seconds by combining filtergrams from both cameras, the front camera and the side camera. Within the 90-second, the front camera takes two sets of Left and Right Circular Polarizations (LCP and RCP) at 6 wavelengths; the side camera takes one set of Linear Polarizations (I+/-Q and I+/-U) at 6 wavelengths. By combining two cameras, one can obtain full Stokes parameters of [I,Q,U,V] at 6 wavelengths in 90 seconds. In norminal Mod-C sequence that HMI currently uses, the front camera takes LCP and RCP at a cadence of 45 seconds, while the side camera takes observation of the full Stokes at a cadence of 135 seconds. Mod-L should be better than Mod-C for providing vector magnetic field data because (1) Mod-L increases cadence of full Stokes observation, which leads to higher temporal resolution of vector magnetic field measurement; (2) decreases noise in vector magnetic field data because it uses more filtergrams to produce [I,Q,U,V]. There are two potential issues in Mod-L that need to be addressed: (1) scaling intensity of the two cameras’ filtergrams; and (2) if current polarization calibration model, which is built for each camera separately, works for the combined data from both cameras. This presentation will address these questions, and further place a discussion here.
- Publication:
-
AAS/Solar Physics Division Abstracts #47
- Pub Date:
- May 2016
- Bibcode:
- 2016SPD....47.0810L