Final report of supplementary comparison AFRIMETS.AUV.A-S1: primary pressure calibration of LS2aP microphones according to IEC 61094-2, over the frequency range 1 Hz to 31.5 kHz.
Abstract
This is the final report of the AFRIMETS.AUV-S1 comparison of the pressure sensitivity, modulus and phase, of LS2aP microphones in the frequency range 1 Hz to 31.5 kHz in accordance with IEC 61094-2. Six national metrology institutes from three different regional metrology organisations participated in the comparison for which two LS2aP microphones were circulated simultaneously to all the participants in a hybrid-star configuration. The comparison reference values were calculated as the weighted mean for modulus and phase for each individual microphone. Main text
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).- Publication:
-