Multichannel Spectrometer of Time Distribution
Abstract
For research and control of characteristics of radiation fluxes, radioactive sources in particular, for example, in paper [1], a spectrometer and methods of data measurement and processing based on the multichannel counter of time intervals of accident events appearance (impulses of particle detector) MC-2A (SPC "ASPECT") were created. The spectrometer has four independent channels of registration of time intervals of impulses appearance and correspondent amplitude and spectrometric channels for control along the energy spectra of the operation stationarity of paths of each of the channels from the detector to the amplifier. The registration of alpha-radiation is carried out by the semiconductor detectors with energy resolution of 16-30 keV. Using a spectrometer there have been taken measurements of oscillations of alpha-radiation 239-Pu flux intensity with a subsequent autocorrelative statistical analysis of the time series of readings.
- Publication:
-
Exotic Nuclei: EXON-2014 - Proceedings of International Symposium. Edited by PENIONZHKEVICH YURI & SOBOLEV YURI. Published by World Scientific Publishing Co. Pte. Ltd
- Pub Date:
- June 2015
- DOI:
- 10.1142/9789814699464_0066
- Bibcode:
- 2015exon.conf..651A