Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors
Abstract
We investigated the performances of two digital mammography systems. The systems use a cesium-iodide (CsI) scintillator and an amorphous selenium (a-Se) photoconductor for X-ray detection and are installed in the same hospital. As physical metrics, we measured the modulationtransfer function (MTF), the noise-power spectrum (NPS), and the detective quantum efficiency (DQE). In addition, we analyzed the contrast-detail performances of the two systems by using a commercial contrast-detail phantom. The CsI-based indirect conversion detector provided better MTF and DQE performances than the a-Se-based direct conversion detector whereas the former provided a poorer NPS performance than the latter. These results are explained by the fact that the CsI-based detector used an MTF restoration preprocessing algorithm. The a-Se-based detector showed better contrast-detail performance than the CsI-based detector. We believe that the highfrequency noise characteristic of a detector is more responsible for the visibility of small details than its spatial-resolution performance.
- Publication:
-
Journal of Korean Physical Society
- Pub Date:
- June 2015
- DOI:
- 10.3938/jkps.66.1926
- Bibcode:
- 2015JKPS...66.1926Y
- Keywords:
-
- Digital mammography;
- Cesium iodide;
- Amorphous selenium;
- MTF;
- NPS;
- DQE;
- Contrast