Improved CRDS δ13C Stability Through New Calibration Application For CO2 And CH4
Abstract
Stable carbon isotope ratio measurements of CO2 and CH4 provide valuable insight into global and regional sources and sinks of the two most important greenhouse gases. Methodologies based on Cavity Ring-Down Spectroscopy (CRDS) have been developed and are capable of delivering δ13C measurements with a precision better than 0.12 permil for CO2 and 0.4 permil for CH4 (1 hour window, 5 minute average). Here we present a method to further improve this measurement stability. We have developed a two-point calibration method which corrects for δ13C drift due to a dependence on carbon species concentration. This method calibrates for both carbon species concentration as well as δ13C. In addition, we further demonstrate that this added stability is especially valuable when using carbon isotope data in linear regression models such as Keeling plots, where even small amounts of error can be magnified to give inconclusive results. Furthermore, we show how this method is used to validate multiple instruments simultaneously and can be used to create the standard samples needed for field calibrations.
- Publication:
-
EGU General Assembly Conference Abstracts
- Pub Date:
- April 2015
- Bibcode:
- 2015EGUGA..17.8039R