Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
Abstract
Currently inorganic scintillator detectors are used in all commercial Time of Flight Positron Emission Tomograph (TOF-PET) devices. The J-PET collaboration investigates a possibility of construction of a PET scanner from plastic scintillators which would allow for single bed imaging of the whole human body. This paper describes a novel method of hit-position reconstruction based on sampled signals and an example of an application of the method for a single module with a 30 cm long plastic strip, read out on both ends by Hamamatsu R4998 photomultipliers. The sampling scheme to generate a vector with samples of a PET event waveform with respect to four user-defined amplitudes is introduced. The experimental setup provides irradiation of a chosen position in the plastic scintillator strip with an annihilation gamma quanta of energy 511 keV. The statistical test for a multivariate normal (MVN) distribution of measured vectors at a given position is developed, and it is shown that signals sampled at four thresholds in a voltage domain are approximately normally distributed variables. With the presented method of a vector analysis made out of waveform samples acquired with four thresholds, we obtain a spatial resolution of about 1 cm and a timing resolution of about 80 ps (σ).
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- November 2014
- DOI:
- 10.1016/j.nima.2014.07.032
- arXiv:
- arXiv:1407.8293
- Bibcode:
- 2014NIMPA.764..186R
- Keywords:
-
- Positron Emission Tomography;
- Plastic scintillators;
- Statistical analysis;
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- doi:10.1016/j.nima.2014.07.032