Characteristics of a CsI(Tl) Compton-suppressed clover detector
Abstract
The performance of a four fold segmented clover detector coupled to a CsI-suppressed shield has been tested using several standard radioactive sources. The summing characteristics, hit patterns and absolute efficiency were measured in both crystal and clover modes. Peak-to-total ratios of 57.3% and 45.6% in suppressed clover mode have been determined for 137Cs and 60Co sources, respectively. A Geant4 simulation of the suppressed clover detector, including the segmentation of the germanium crystals, has been created, and the simulation results agreed very well with the experimental data.