PILOT STUDY: Report on the CCPR Pilot Comparison: Spectral Responsivity 10 nm to 20 nm
Abstract
The CCPR Pilot Comparison on spectral responsivity in the 10 nm to 20 nm spectral range was carried out within the framework of the CIPM Mutual Recognition Arrangement by three laboratories: PTB (Germany), NIST (USA), and NMIJ/AIST (Japan) with PTB acting as the central and reporting laboratory. All participating laboratories used monochromatized synchrotron radiation. PTB and NIST used a cryogenic radiometer as the primary standard detector and NMIJ, an ionization chamber with extrapolation by a wavelength-independent detector.
The aim of the pilot comparison was to check the accuracy of the radiometric scale of spectral responsivity in the short wavelength EUV spectral range which has recently gained in technological importance. The wavelengths of measurement were from 11.5 nm to 20 nm in 0.5 nm steps and additionally 12.2 nm. The comparison was carried out through the calibration of a group of transfer standard detectors. Two sets of three diodes of types AXUV and SXUV from International Radiation Detectors, Inc. were used for the comparison. The comparison had the form of a star comparison: Pilot-lab A-pilot-lab B-pilot, PTB acting as the pilot laboratory. All results were communicated directly to the pilot laboratory. The report describes in detail the measurements made at PTB and summarizes the reports submitted by the participants. Measurements carried out by the pilot laboratory before and after the circulation of the detectors proved that the stability of the detectors was sufficient for the comparison. For the type AXUV detectors, however, changes in their responsivity contributed to the uncertainty of the comparison. Measurement results from participants and their associated uncertainties were analyzed in this report according to the Guidelines for CCPR Comparison Report Preparation. The uncertainty contributions were separated, as to whether they are wavelength dependent or not. All bilateral DoE are well within the respective k = 2 expanded uncertainty ranges for all wavelengths. The separated non-wavelength-dependent and wavelength-dependent uncertainty contributions are consistent with the respective wavelength-averaged DoE and wavelength-dependent variations of all bilateral DoE. Main text. To reach the main text of this paper, click on Final Report. The final report has been peer-reviewed and approved for publication by the CCPR WG-KC- Publication:
-
Metrologia
- Pub Date:
- January 2010
- DOI:
- 10.1088/0026-1394/47/1A/02001
- Bibcode:
- 2010Metro..47.2001S