Development of an in situ gas chromatograph - mass selective detector for the purpose of studying long-range pollution transport from Asia
Abstract
In September 2002, we began work on the development and construction of an automated, 3 channel gas chromatograph (GC) with a mass selective detector (MSD) and two electron capture detectors (ECDs). The instrument will be placed at a Pacific CMDL station (Mauna Loa or Trinidad Head), and will make continuous measurements of a variety of chemical species with a wide range of lifetimes, including chloroflourocarbons (CFCs), hydrochlorofluorocarbons (HCFCs), hydrofluorocarbons (HFCs), methyl halides, non-methane hydrocarbons (NMHCs), peroxyacetyl nitrate (PAN), nitrous oxide, and sulfur hexafluoride. The primary goal is to characterize the episodic long-range transport of pollution from Asia. We have completed development of a cryogenic trapping system for pre-concentrating samples for the MSD channel. We present a comparison and characterization of two types of traps, one with a one-inch section of Porapak Q as the adsorbent material, and one with 6 mg of Carboxen 1003 and 4 mg of Carboxen 1000. We discuss the suitability of each of these traps for our intended research. We also present our progress on other aspects of the measurement system, including a dynamic dilution system for calibration of PAN measurements on one of the ECD channels, and a discussion of the science issues involved in choosing the deployment location.