Generalized method for the calibration of 4-terminal-pair type digital impedance meters
Abstract
A calibration procedure for digital impedance meters is described and its use is illustrated. The calibration is accomplished through the use of impedance standards which relate instrument readings to the values of the standards through a known functional relationship. The calibration procedure described estimates the parameters associated with the functional relationship and requires the use of a computer. Calibration is accomplished at the reference plane of the impedance standards and any adapter required to connect the standards to the instrument is assumed to be an integral part of the impedance meter.
- Publication:
-
NASA STI/Recon Technical Report N
- Pub Date:
- August 1984
- Bibcode:
- 1984STIN...8520318J
- Keywords:
-
- Calibrating;
- Impedance Measurement;
- Computer Programs;
- Fortran;
- Least Squares Method;
- Reflectance;
- Instrumentation and Photography