A new particle-induced X-ray emission set-up for laterally resolved analysis over wide areas
Abstract
The recently installed and unique PIXE (particle-induced X-ray emission) set-up at the Helmholtz-Zentrum Dresden-Rossendorf (HZDR) is mainly dedicated to applications for a detailed overview of elemental composition over large sample areas within a short time even at trace level. The so-called High-Speed-PIXE (HS-PIXE), a combination of a pnCCD-based pixel-detector with polycapillary X-ray optics, offers simultaneous imaging of sample areas up to 12 × 12 mm2 with a lateral resolution better than 100 μm. Each of the 264 × 264 individual pixels detects X-ray photons in an energy range from 2 keV to 20 keV with an energy resolution of 152 eV (@Mn-Kα). A high precision sample manipulator offers the inspection of areas up to 250 × 250 mm2. During first experiments the determined resolution is (76 ± 23) μm using a sample of well-known sharp-edged chromium patterns. Trace element analysis has been performed using a geological sample, a tin ore, with an average Ta-concentration below 0.1 at.%. Fine-zoned structures became visible in the Ta-Lα intensity map within only 45 min. The High-Speed-PIXE closes a gap in the analytical process flow chain especially for geoanalytical characterisations. It is a unique and fast detection system to identify areas of interest in comparably short time at large-area scale for further analysis.
- Publication:
-
Nuclear Instruments and Methods in Physics Research B
- Pub Date:
- June 2016
- DOI:
- 10.1016/j.nimb.2016.03.032
- Bibcode:
- 2016NIMPB.377...17H
- Keywords:
-
- PIXE;
- Lateral resolution;
- Imaging;
- Geometallurgy