A High-Precision Multi-Channel TAC and QAC Module for the Neutron Detection Wall
Abstract
A single width NIM module that includes eight channels TAC (time-to-amplitude converter) and QAC (charge-to-amplitude converter) is introduced in the paper, which is designed for the large neutron wall detector to measure charge (energy) and time interval simultaneously [1]. The module mainly adopts a high precision gated integral circuit to realize TAC and QAC. The input range of TAC is from 30 ns to 1 us, and the input range of QAC is from 40 pC to 600 pC. The linearity error of TAC is lower than 1.28 %, and the time resolution of TAC is less than 0.871 %. The linearity error of QAC is lower than 0.81 %, and the resolution of QAC is better than 0.936 %.