a Solution for Dosimetry and Quality Assurance in Imrt and Hadrontherapy:. the Pixel Ionization Chamber
Abstract
The new radiotherapy techniques require new detectors to monitor and measure the clinical field. The Intensity Modulated Radiation Therapy (IMRT) techniques like step and shoot, sliding window, dynamic wedge or scanning beam add the time variable to the treatment field. In this case the water phantom with a single ionization chamber moving inside the field needs very long measurement time. Linear arrays of ionization chambers or diodes measure the field only along a line. 2D detectors like radiographic or gafchromic film are not suitable to be used as on line detectors. We have developed, built and tested an ionization chamber segmented in pixels that measure the dose in a plane at several points. Every channel has a dedicated electronic chain that digitizes the collected charge and data from all the channels are sent to the computer that performs the data acquisition. One read out cycle is very fast allowing to measure in real time the fluency and the shape of the field. The chamber can be used in two different ways, as monitor chamber and as relative dosemeter. A description of the detector, the electronics, and test results with both photon and hadron beams will be reported.