Gamma ray spectroscopy with scintillation light in liquid xenon
Abstract
Scintillation light from gamma ray irradiation in liquid xenon is detected by two Hamamatsu R9288 photomultiplier tubes (PMTs) immersed in the liquid. UV light reflector material, PTFE, is used to optimize the light collection efficiency. The detector gives a high light yield of 6 photoelectron per keV (pe/keV), which allows efficient detection of the 122 keV γ-ray line from 57Co, with a measured energy resolution of (8.8±0.6)% (σ). The best achievable energy resolution, by removing the instrumental fluctuations, from liquid xenon scintillation light is estimated to be around 6-8% (σ) for γ-ray with energy between 662 keV and 122 keV.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- September 2006
- DOI:
- arXiv:
- arXiv:physics/0608034
- Bibcode:
- 2006JInst...1.9004N
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- JINST 1 (2006) P09004