Leakage tests for 241Am solid sources used for liquid xenon detector monitoring
Abstract
Experimental tests are described to check solid 241Am radioactive sources used to monitor operation of liquid xenon detectors. In particular, radioactive leakage was tested in extreme temperature conditions following immersion in liquid nitrogen for different time periods. No radioactivity loss was detected from the sources considered. The paper describes the source characteristics, test methods and results obtained.