Self-cooling cryogenic microcalorimeters made of SINIS junctions
Abstract
High-quality low-leakage SINIS devices made of Al-AlMn-Al layers were fabricated for energy dispersive single photon detection. Information on different heat flow channels was extracted from the measured dynamics of detector signals due to X-ray events. At the optimum operation point, the extracted effective electron temperature decreased from 88 down to 43 mK due to self-cooling, roughly doubling the detector sensitivity.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- April 2006
- DOI:
- 10.1016/j.nima.2005.12.091
- arXiv:
- arXiv:physics/0507003
- Bibcode:
- 2006NIMPA.559..636F
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 3 pages, 3 figures