Novel Data Acquisition System for Silicon Tracking Detectors
Abstract
We have developed a novel data acquisition system for measuring tracking parameters of a silicon detector in a particle beam. The system is based on a commercial Analog-to-Digital VME module and a PC Linux based Data Acquisition System. This DAQ is realized with C++ code using object-oriented techniques. Track parameters for the beam particles were reconstructed using off-line analysis code and automatic detector position alignment algorithm. The new DAQ was used to test novel Czochralski type silicon detectors. The important silicon detector parameters, including signal size distributions and signal to noise distributions, were successfully extracted from the detector under study. The efficiency of the detector was measured to be 95 %, the resolution about 10 micrometers, and the signal to noise ratio about 10.
- Publication:
-
arXiv e-prints
- Pub Date:
- June 2003
- DOI:
- arXiv:
- arXiv:physics/0306015
- Bibcode:
- 2003physics...6015W
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- Talk from the 2003 Computing in High Energy and Nuclear Physics (CHEP03), La Jolla, Ca, USA, March 2003, 6 pages, LaTeX, 5 eps figures. PSN TUGP006