Intense pulsed helium droplet beams
Abstract
Pulsed (30-100 μs) nozzle beams have been used to generate helium droplets (<N>=104-105). The dependence of the beam intensity and the mean droplet size on the source stagnation pressure and temperature are studied via mass spectroscopy and laser induced fluorescence of embedded phthalocyanine molecules. In comparison to a cw beam the pulsed source for the same pressure and temperature has a factor of 100 higher flux and the droplet sizes are 1 order of a magnitude larger.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- October 2002
- DOI:
- 10.1063/1.1505661
- arXiv:
- arXiv:physics/0206041
- Bibcode:
- 2002RScI...73.3600S
- Keywords:
-
- 67.40.-w;
- 47.55.Dz;
- 07.20.Mc;
- 36.40.Mr;
- 39.10.+j;
- 07.77.Gx;
- Boson degeneracy and superfluidity of <sup>4</sup>He;
- Cryogenics;
- refrigerators low-temperature detectors and other low-temperature equipment;
- Spectroscopy and geometrical structure of clusters;
- Atomic and molecular beam sources and techniques;
- Atomic and molecular beam sources and detectors;
- Physics - Atomic and Molecular Clusters