Effect of tip geometry on contrast and spatial resolution of the near-field microwave microscope
Abstract
The near-field scanning microwave microscope (NSMM) can quantitatively image materials properties at length scales far shorter than the free space wavelength (λ). Here we report a study of the effect of tip geometry on the NSMM signals. This particular NSMM utilizes scanning tunneling microscopy (STM) for distance-following control. We systematically examined many commercially available STM tips and found them to have a conical structure on the macroscopic scale, with an embedded sphere (of radius rsphere) at the apex of the tip. The rsphere values used in the study ranged from 0.1to12.6μm. Tips with larger rsphere show good signal contrast [as measured by the frequency shift (Δf) signal between tunneling height and 2μm away from the sample] with NSMM. For example, the tips with rsphere=8μm give signal contrast of 1000kHz compared to 85kHz with a tip of rsphere=0.55μm. However, large rsphere tips distort the topographic features acquired through STM. A theoretical model is used to understand the tip-to-sample interaction. The model quantitatively explains the measured change in quality factor (Q) as a function of height over bulk copper and silicon samples.