Establishing Visible Interferometer System Responses: Resolved and Unresolved Calibrators
Abstract
This investigation examines the propagation of errors through the uniform disk visibility function. The implications of those errors on measurements of absolute visibility through optical and near-infrared interferometers are considered within the context of using calibration stars to establish system visibilities for these instruments. We suggest a simple ratio test to establish empirically whether or not the measured visibilities produced by such an instrument are relative (errors dominated by calibrator angular size prediction error) or absolute (errors dominated by measurement error).
- Publication:
-
Publications of the Astronomical Society of the Pacific
- Pub Date:
- November 2005
- DOI:
- arXiv:
- arXiv:astro-ph/0508266
- Bibcode:
- 2005PASP..117.1263V
- Keywords:
-
- Instrumentation: Interferometers;
- Methods: Analytical;
- Techniques: Interferometric;
- Stars: Fundamental Parameters;
- Astrophysics
- E-Print:
- 20 pages, 7 figures, to be published in the PASP