Drift time calibration of the ultra-Low material budget GEM-based TPC for MIXE
Abstract
Muon-Induced X-ray Emission (MIXE) is a non-destructive analytical technique that leverages negative muons to probe elemental and isotopic compositions by detecting characteristic muonic X-rays emitted during atomic cascades and gamma rays from nuclear capture processes. By controlling the muon beam momentum, MIXE enables depth-resolved analysis, spanning microns to centimeters, making it ideal for studying compositional variations in fragile, valuable, or operando samples. To enhance its capabilities, we integrated a twin Time Projection Chamber (TPC) tracker with Gas Electron Multiplier (GEM) amplification stages, allowing precise measurement of muon trajectories. A custom-built fiber detector with scintillating fibers and a Silicon Photomultiplier (SiPM) provided sub-percent-level accuracy in drift velocity calibration, essential for accurate spatial reconstruction. This advanced setup correlates muon stopping points with X-ray emissions, paving the way towards element-sensitive imaging and establishing MIXE as a unique tool for high-resolution, depth-specific elemental analysis across diverse scientific applications.
- Publication:
-
arXiv e-prints
- Pub Date:
- January 2025
- arXiv:
- arXiv:2501.10249
- Bibcode:
- 2025arXiv250110249Z
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 4 pages, 3 figures, conference proceeding MPGD2024