Silicon drift detectors for the Spectroscopy Focusing Array of eXTP
Abstract
We present a silicon drift detector (SDD) system for the spectroscopy focusing array (SFA) of the enhanced X-ray timing and polarimetry (eXTP) mission. The SFA focuses on fast timing (time resolution below 10 {\mu}s) and good spectroscopy capabilities (energy resolution better than 180 eV @ 6 keV). The sensor, consisting of 19 hexagonally shaped pixels with a total sensitive area of ${5.05}\, cm^{2}$, is connected to three high time resolution spectroscopy (HTRS) ASICs, allowing a fast readout of the detector signals. The detector works in a Charge- Sensitive Amplifier configuration. We assembled a prototype detector module and present here its mechanical design, describe the used sensor, and report about its performance.
- Publication:
-
arXiv e-prints
- Pub Date:
- July 2024
- DOI:
- arXiv:
- arXiv:2407.09185
- Bibcode:
- 2024arXiv240709185A
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics
- E-Print:
- 9 pages, 5 figures