Characterization of a radiation detector based on opaque water-based liquid scintillator
Abstract
We present the characterization of a novel radiation detector based on an opaque water-based liquid scintillator. Opaque scintillators, also known as LiquidO, are made to be highly scattering, such that the scintillation light is effectively confined, and read out through wavelength-shifting fibers. The 1-liter, 32-channel prototype demonstrates the capability for both spectroscopy and topological reconstruction of point-like events. The design, construction, and evaluation of the detector are described, including modeling of the scintillation liquid optical properties and the detector's response to gamma rays of several energies. A mean position reconstruction error of 4.4 mm for 1.6 MeV-equivalent events and 7.4 mm for 0.8 MeV-equivalent events is demonstrated using a simple reconstruction approach analogous to center-of-mass.
- Publication:
-
arXiv e-prints
- Pub Date:
- June 2024
- DOI:
- 10.48550/arXiv.2406.13054
- arXiv:
- arXiv:2406.13054
- Bibcode:
- 2024arXiv240613054L
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 20 pages, 14 figures