X-ray Interferometry Using a Modulated Phase Grating: Theory and Experiments
Abstract
X-ray grating interferometry allows for the simultaneous acquisition of attenuation, differential-phase contrast, and dark-field images, resulting from X-ray attenuation, refraction, and small-angle scattering, respectively. The modulated phase grating (MPG) interferometer is a recently developed grating interferometry system capable of generating a directly resolvable interference pattern using a relatively large period grating envelope function that is sampled at a pitch that allows for X-ray spatial coherence using a microfocus X-ray source or by use of a source G0 grating that follows the Lau condition. We present the theory of the MPG interferometry system for a 2-dimensional staggered grating, derived using Fourier optics, and we compare the theoretical predictions with experiments we have performed with a microfocus X-ray system at Pennington Biomedical Research Center, LSU. The theoretical and experimental fringe visibility is evaluated as a function of grating-to-detector distance. Quantitative experiments are performed with porous carbon and alumina samples, and qualitative analysis of attenuation and dark-field images of a dried anchovy are shown.
- Publication:
-
arXiv e-prints
- Pub Date:
- April 2024
- DOI:
- 10.48550/arXiv.2404.14584
- arXiv:
- arXiv:2404.14584
- Bibcode:
- 2024arXiv240414584M
- Keywords:
-
- Physics - Optics;
- Physics - Instrumentation and Detectors;
- Physics - Medical Physics
- E-Print:
- 22 pages, 13 figures. This is a pre-print