Radon concentration measurement with a high-sensitivity radon detector at the Yemilab
Abstract
The radiation emitted from radon is a critical background in rare event search experiments conducted at the Yemi Underground Laboratory (Yemilab) in Jeongseon, Korea. A Radon Reduction System (RRS) has been developed and installed in Yemilab to reduce radon concentration in the air. The RRS primarily provides a purified air of 50 m3/h to the cleanroom used to assemble crystal detectors in the AMoRE, a neutrinoless double beta decay search experiment. RRS can reduce the radon level by a factor of 300, so a high-sensitivity radon detector was required. A highly sensitive radon detector was constructed using a 70 L chamber with a large PIN photodiode to measure radon concentration in the purified air. The radon detector shows an excellent resolution of 72 keV (FWHM) for 6.003 MeV alphas from 218Po decay and a sensitivity down to 23.8 ± 2.1 mBq/m3 with a boil-off N2 gas sample. The radon concentration level from the RRS measured by the radon detector was below 0.29 Bq/m3 with a reduction factor of about 300.