Microcalorimeters that leverage microwave kinetic inductance detectors to read out phonon signals in the particle-absorbing target, referred to as kinetic inductance phonon-mediated (KIPM) detectors, offer an attractive detector architecture to probe dark matter (DM) down to the fermionic thermal relic mass limit. A prototype KIPM detector featuring a single aluminum resonator patterned onto a 1-gram silicon substrate was operated in the Northwestern EXerimental Underground Site (NEXUS) low-background facility at Fermilab for characterization and evaluation of this detector architecture's efficacy for a DM search. An energy calibration was performed by exposing the bare substrate to a pulsed source of 470-nm photons, resulting in a baseline resolution on the energy absorbed by the phonon sensor of