MIDSX: A Monte Carlo Interaction and Dosimetry Simulation of X-rays
Abstract
Computed Tomography (CT) imaging, while essential for diagnostics, exposes patients to ionizing radiation. To accurately quantify radiation dosage, this study introduces MIDSX, a specialized open-source Monte Carlo (MC) photon transport code system for medical imaging. Unlike general purpose particle transport MC systems, MIDSX is tailored for x-ray transport, offering more streamlined implementation. Results from validation simulations show MIDSX's results for specific cases agree to within 0.39% of the mean of established reference data. However, discrepancies in body energy deposition measurements, reaching up to a 6.3% mean percent error, indicate areas for further refinement in the system.
- Publication:
-
arXiv e-prints
- Pub Date:
- November 2023
- DOI:
- 10.48550/arXiv.2311.16873
- arXiv:
- arXiv:2311.16873
- Bibcode:
- 2023arXiv231116873M
- Keywords:
-
- Physics - Medical Physics
- E-Print:
- 4 pages, 3 figures, submitted to the Journal of Undergraduate Reports in Physics (JURP)