A versatile and narrow linewidth infrared radiation source for ro-vibration state-selected preparation of molecules in molecular beams
Abstract
We describe the design and characterization of a versatile pulsed (5 ns, 10 Hz repetition rate) optical parametric oscillator and amplifier system capable of generating single longitudinal mode, narrow linewidth (0.01 cm-1) radiations in the wavelength range of 680-870 nm and 1380-4650 nm. Using a combination of power-normalized photoacoustic signal and a Fizeau interferometer-based wavemeter, we are able to actively stabilize the output wavenumber to within 0.005 cm-1 (3σ) over a timescale longer than 1000 s. We demonstrate an application of this system by performing ro-vibration state-selected preparation of CO in the v = 2 state, via direct overtone excitation (v = 0 → 2 at 2346 nm) and subsequent state-selected detection in an internally cold molecular beam.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- November 2023
- DOI:
- 10.1063/5.0157912
- arXiv:
- arXiv:2305.19792
- Bibcode:
- 2023RScI...94k3001K
- Keywords:
-
- Physics - Optics;
- Physics - Chemical Physics;
- Physics - Instrumentation and Detectors
- E-Print:
- 8 pages, 8 figures + supplementary information