Voltage scanning and technical upgrades at the Collinear Resonance Ionization Spectroscopy experiment
Abstract
To optimize the performance of the Collinear Resonance Ionization Spectroscopy (CRIS) experiment at CERN-ISOLDE, technical upgrades are continuously introduced, aiming to enhance its sensitivity, precision, stability, and efficiency. Recently, a voltage-scanning setup was developed and commissioned at CRIS, which improved the scanning speed by a factor of three as compared to the current laser-frequency scanning approach. This leads to faster measurements of the hyperfine structure for systems with high yields (more than a few thousand ions per second). Additionally, several beamline sections have been redesigned and manufactured, including a new field-ionization unit, an electrostatic bend with a larger deflection angle, and improved ion optics. The beamline upgrades are expected to yield an improvement of at least a factor of 5 in the signal-to-noise ratio by avoiding the use of high-power lasers (which yield non-resonantly produced ions) and providing time-of-flight separation between the resonant ions and the collisional background.
- Publication:
-
Nuclear Instruments and Methods in Physics Research B
- Pub Date:
- August 2023
- DOI:
- arXiv:
- arXiv:2303.15312
- Bibcode:
- 2023NIMPB.541...86A
- Keywords:
-
- CRIS;
- Collinear resonance ionization spectroscopy;
- Laser spectroscopy;
- Voltage scanning;
- ISOLDE;
- Physics - Instrumentation and Detectors;
- Nuclear Experiment
- E-Print:
- 10 pages. Under review at NIM B as part of the proceedings of EMIS 2022 at RAON, South Korea