Evaluation of the Constant Fraction Time-Over-Threshold (CF-TOT) method for neutron-gamma pulse shape discrimination
Abstract
The use of Time-over-Threshold (TOT) for the discrimination between fast neutrons and gamma-rays is advantageous when large number of detection channels are required due to the simplicity of its implementation. However, the results obtained using the standard, Constant Threshold TOT (CT-TOT) are usually inferior to those obtained using other pulse shape discrimination (PSD) methods, such as Charge Comparison or Zero-Crossing approaches, especially for low amplitude neutron/gamma-ray pulses. We evaluate another TOT approach for fast neutron/gamma-ray PSD using Constant-Fraction Time-over-Threshold (CF-TOT) pulse shape analysis. The CT-TOT and CF-TOT methods were compared quantitatively using digitized waveforms from a liquid scintillator coupled to a photomultiplier tube as well as from a stilbene scintillator coupled to a photomultiplier tube and a silicon photomultiplier. The quality of CF-TOT neutron/gamma-ray discrimination was evaluated using Receiver Operator Characteristics curves and the results obtained with this approach were compared to that of the standard CT-TOT method. The CF-TOT PSD method results in > 99.9% rejection of gamma-rays with > 80% neutron acceptance, much better than CT-TOT.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- May 2022
- DOI:
- 10.1088/1748-0221/17/05/P05028
- arXiv:
- arXiv:2112.02347
- Bibcode:
- 2022JInst..17P5028R
- Keywords:
-
- Data acquisition concepts;
- Neutron detectors (cold;
- thermal;
- fast neutrons);
- Particle identification methods;
- Scintillators and scintillating fibres and light guides;
- Physics - Instrumentation and Detectors
- E-Print:
- Accepted for publication in JINST