Sensitivity analysis towards trace-uranium detection with γ-γ coincidence NAA
Abstract
We present an improved approach for detection of trace amounts of 238U by means of neutron activation analysis (NAA). The analysis enhancement is obtained by utilizing γ-γ coincidence counting. An empirical method for evaluating the sensitivity gain in the presence of a large source-related background is presented. A comparison of detection limits is made between two counting schemes; namely, counting single gammas using one HPGe detector versus counting coincident gammas using two HPGe detectors placed face-to-face. A data-validated radiation transport model, created in GEANT4 for a single HPGe detector, is extended to handle the two-detector setup. In this counting scheme, a 238U-detection sensitivity enhancement of about a factor of 8 is predicted for a sample of Saint-Gobain G3 Sapphire when compared to the simpler singles counting approach.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- October 2021
- DOI:
- 10.1088/1748-0221/16/10/P10007
- arXiv:
- arXiv:2108.07841
- Bibcode:
- 2021JInst..16P0007T
- Keywords:
-
- Inspection with neutrons;
- Detector modelling and simulations I (interaction of radiation with matter;
- interaction of photons with matter;
- interaction of hadrons with matter;
- etc);
- Gamma detectors (scintillators;
- CZT;
- HPGe;
- HgI etc);
- Physics - Instrumentation and Detectors;
- Nuclear Experiment
- E-Print:
- JINST 16 P10007 (2021)