Strong negative electrothermal feedback in thermal kinetic inductance detectors
Abstract
We demonstrate strong negative electrothermal feedback accelerating and linearizing the response of a thermal kinetic inductance detector (TKID). TKIDs are a proposed highly multiplexable replacement to transition-edge sensors and measure power through the temperature-dependent resonant frequency of a superconducting microresonator bolometer. At high readout probe power and probe frequency detuned from the TKID resonant frequency, we observe electrothermal feedback loop gain up to L ≈ 16 through measuring the reduction of settling time. We also show that the detector response has no detectable non-linearity over a 38% range of incident power and that the noise-equivalent power is below the design photon noise.
- Publication:
-
Journal of Applied Physics
- Pub Date:
- September 2021
- DOI:
- 10.1063/5.0064723
- arXiv:
- arXiv:2107.12493
- Bibcode:
- 2021JAP...130l4503A
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Astrophysics - Instrumentation and Methods for Astrophysics;
- Condensed Matter - Superconductivity
- E-Print:
- 8 pages, 9 figures, The following article has been submitted to the Journal of Applied Physics. After it is published, it will be found at https://publishing.aip.org/resources/librarians/products/journals/