A femtosecond magnetic circular dichroism spectrometer
Abstract
We describe the development of a broadband magneto-optical spectrometer with femtosecond temporal resolution. The absorption spectrometer is based on a white-light supercontinuum (∼320 to 750 nm) using shot-to-shot temporal and spectral referencing at 1 kHz. Static and transient absorption spectra using circularly polarized light are collected in a magnetic field. The difference spectra with respect to the external field direction give the static and transient magneto-optical Faraday rotation (magnetic optical rotary dispersion) and ellipticity (magnetic circular dichroism) spectra. An achromatic quarter-wave plate is used, and the impact of the deviation from ideal retardance on the spectra is discussed. Results from solution-based and thin-film samples are used to demonstrate the performance and wide applicability of the instrument. The sensitivities for the static and time-resolved data were found to be 5 and 0.4 mdeg, respectively. The method presents a simple way to measure magneto-optical spectra using a transient absorption spectrometer and an electromagnet.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- November 2021
- DOI:
- 10.1063/5.0064460
- arXiv:
- arXiv:2107.10729
- Bibcode:
- 2021RScI...92k3001S
- Keywords:
-
- Physics - Optics;
- Condensed Matter - Materials Science
- E-Print:
- The following article has been submitted to Review of Scientific Instruments