In the framework of rare event searches, the identification of radioactive contaminants in ultra-pure samples is a challenging task, because the signal is often at the same level of the instrumental background. This is a rather common situation for α -spectrometers and other detectors used for low-activity measurements. In order to obtain the target sensitivity without extending the data taking live-time, analysis strategies that highlight the presence of the signal sought should be developed. In this paper, we show how to improve the contaminant tagging capability relying on the time-correlation of radioactive decay sequences. We validate the proposed technique by measuring the impurity level of both contaminated and ultra-pure copper samples, demonstrating the potential of this analysis tool in disentangling different background sources and providing an effective way to mitigate their impact in rare event searches.
- Publication:
-
European Physical Journal C
- Pub Date:
- November 2021
- DOI:
- 10.1140/epjc/s10052-021-09759-5
- arXiv:
- arXiv:2105.03140
- Bibcode:
- 2021EPJC...81..971B
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Nuclear Experiment
- E-Print: