Large-Area SiPM Pixels (LASiPs): a cost-effective solution towards compact large SPECT cameras
Abstract
Single Photon Emission Computed Tomography (SPECT) scanners based on photomultiplier tubes (PMTs) are still largely employed in the clinical environment. A standard camera for full-body SPECT employs $\sim50$-100 PMTs of 4-8~cm diameter and is shielded by a thick layer of lead, becoming a heavy and bulky system that can weight a few hundred kilograms. The volume, weight and cost of a camera can be significantly reduced if the PMTs are replaced by silicon photomultipliers (SiPMs). The main obstacle to use SiPMs in full-body SPECT is the limited size of their sensitive area. A few thousand channels would be needed to fill a camera if using the largest commercially-available SiPMs of 6$\times$6~mm$^2$. As a solution, we propose to use Large-Area SiPM Pixels (LASiPs), built by summing individual currents of several SiPMs into a single output. We developed a LASiP prototype that has a sensitive area 8 times larger than a 6$\times$6~mm$^2$ SiPM. We built a proof-of-concept micro-camera consisting of a 40$\times$40$\times$8~mm$^3$ NaI(Tl) crystal coupled to 4 LASiPs. We evaluated its performance in a central region of $15\times15$~mm$^2$, where we were able to reconstruct images of a $^{99m}$Tc capillary with an intrinsic spatial resolution of $\sim2$~mm and an energy resolution of $\sim11.6$\% at 140 keV. We used these measurements to validate Geant4 simulations of the system. This can be extended to simulate a larger camera with more and larger pixels, which could be used to optimize the implementation of LASiPs in large SPECT cameras. We provide some guidelines towards this implementation.
- Publication:
-
arXiv e-prints
- Pub Date:
- February 2021
- DOI:
- 10.48550/arXiv.2102.13402
- arXiv:
- arXiv:2102.13402
- Bibcode:
- 2021arXiv210213402G
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Physics - Applied Physics;
- Physics - Medical Physics
- E-Print:
- Corresponding author: D. Guberman (daniel.guberman@pi.infn.it). Accepted for publication in Physica Medica. 25 pages, 15 figures