Abstract
A microwave superconducting quantum interference device multiplexer has been optimized for reading out large arrays of superconducting transition-edge sensor (TES) bolometers. We present the scalable cryogenic multiplexer chip design that may be used to construct an 1820-channel multiplexer for the 4-8 GHz rf band. The key metrics of yield, sensitivity, and crosstalk are determined through measurements of 455 readout channels, which span 4-5 GHz. The median white-noise level is 45 pA/ √{ Hz } , evaluated at 2 Hz, with a 1/f knee ≤ 20 mHz after common-mode subtraction. The white-noise level decreases the sensitivity of a TES bolometer optimized for detection of the cosmic microwave background at 150 GHz by only 3%. The measured crosstalk between any channel pair is ≤ 0.3%.
- Publication:
-
Applied Physics Letters
- Pub Date:
- February 2021
- DOI:
- 10.1063/5.0033416
- arXiv:
- arXiv:2010.07998
- Bibcode:
- 2021ApPhL.118f2601D
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics
- E-Print:
- 6 pages, 5 figures