Charge collection efficiency in back-illuminated Charge-Coupled Devices
Abstract
Low noise CCDs fully-depleted up to 675 micrometers have been identified as a unique tool for Dark Matter searches and low energy neutrino physics. The charge collection efficiency (CCE) for these detectors is a critical parameter for the performance of future experiments. We present here a new technique to characterize CCE in back-illuminated CCDs based on soft X-rays. This technique is used to characterize two different detector designs. The results demonstrate the importance of the backside processing for detection near threshold, showing that a recombination layer of a few microns significantly distorts the low energy spectrum. The studies demonstrate that the region of partial charge collection can be reduced to less than 1 micrometer thickness with adequate backside processing.
- Publication:
-
arXiv e-prints
- Pub Date:
- July 2020
- DOI:
- 10.48550/arXiv.2007.04201
- arXiv:
- arXiv:2007.04201
- Bibcode:
- 2020arXiv200704201F
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- Phys. Rev. Applied 15, 064026 (2021)