Characterization of the soft X-ray spectrometer PEAXIS at BESSY II
Abstract
The performance of the recently commissioned spectrometer PEAXIS for resonant inelastic soft X-ray scattering and X-ray photoelectron spectroscopy and its hosting beamline U41-PEAXIS at the BESSY II synchrotron are experimentally characterized. With the design of the instrument, a 5 m-long RIXS spectrometer arm that can be continuously rotated by 106° and its sample environment for single crystalline materials covering a broad temperature range 10 K-1000 K, PEAXIS targets fundamental science in solid state physics where access to a range of wavevector transfers is imperative.
- Publication:
-
Journal of Synchrotron Radiation
- Pub Date:
- January 2020
- DOI:
- 10.1107/S1600577519014887
- arXiv:
- arXiv:1906.09455
- Bibcode:
- 2020JSynR..27..238S
- Keywords:
-
- resonant inelastic X-ray scattering;
- X-ray photoelectron spectroscopy;
- soft X-ray spectroscopy;
- soft X-ray beamline;
- X-ray emission;
- X-ray absorption;
- BESSY II;
- Physics - Instrumentation and Detectors;
- Condensed Matter - Strongly Correlated Electrons
- E-Print:
- 12 pages, 14 figures