Alignment-free characterization of polarizing beamsplitters
Abstract
Traditional methods for measurement of Polarizing Beamsplitter (PBS) parameters, especially the extinction ratio, require highly polarized light sources, alignment procedures and/or experimental parameters that change over time, such as polarization rotations. In this work, a new method is presented, which employs unpolarized light and a Faraday Mirror. It is shown that precise extinction ratio and insertion loss values can be achieved in three single-sweep measurements, without any alignment requirements or time-varying signals of any kind.