Segmented YSO scintillation detectors as a new β-implant detection tool for decay spectroscopy in fragmentation facilities
Abstract
A newly developed segmented YSO scintillator detector was implemented for the first time at the RI-beam Factory at RIKEN Nishina Center as an implantation-decay counter. The results from the experiment demonstrate that the detector is a viable alternative to conventional silicon-strip detectors with its good timing resolution and high detection efficiency for β particles. A Position-Sensitive Photo-Multiplier Tube (PSPMT) is coupled with a 48 × 48 segmented YSO crystal. To demonstrate its capabilities, a known short-lived isomer in 76Ni and the β decay of 74Co were measured by implanting those ions into the YSO detector. The half-lives and γ-rays observed in this work are consistent with the known values. The β-ray detection efficiency is more than 80 % for the decay of 74Co.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- September 2019
- DOI:
- 10.1016/j.nima.2019.05.026
- arXiv:
- arXiv:1903.03670
- Bibcode:
- 2019NIMPA.937...93Y
- Keywords:
-
- Beta-decay;
- Implant-beta detector;
- Radioactive isotopes;
- Fragmentation;
- Physics - Instrumentation and Detectors
- E-Print:
- 17 pages, 6 figures, accepted to NIMA on May 7, 2019