Reduced damage in electron microscopy by using interaction-free measurement and conditional re-illumination
Abstract
Interaction-free measurement (IFM) has been proposed as a means of high-resolution, low-damage imaging of radiation-sensitive samples, such as biomolecules and proteins. The basic setup for IFM is a Mach-Zehnder interferometer, and recent progress in nanofabricated electron diffraction gratings has made it possible to incorporate a Mach-Zehnder interferometer in a transmission-electron microscope (TEM). Therefore, the limits of performance of IFM with such an interferometer and a shot-noise limited electron source (such as that in a TEM) are of interest. In this work, we compared the error probability and sample damage for ideal IFM and classical imaging schemes, through theoretical analysis and numerical simulation. We considered a sample that is either completely transparent or completely opaque at each pixel. In our analysis, we also evaluated the impact of an additional detector for scattered electrons. The additional detector resulted in reduction of error by up to an order of magnitude, for both IFM and classical schemes. We also investigated a sample re-illumination scheme based on updating priors after each round of illumination and found that this scheme further reduced error by a factor of two. Implementation of these methods is likely achievable with existing instrumentation and would result in improved resolution in low-dose electron microscopy.
- Publication:
-
arXiv e-prints
- Pub Date:
- January 2019
- DOI:
- 10.48550/arXiv.1901.09702
- arXiv:
- arXiv:1901.09702
- Bibcode:
- 2019arXiv190109702A
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 16 pages