Measurement of ^{144}Ce - ^{144}Pr beta-spectra with Si(Li) detectors for the purpose of determining the spectra of electron antineutrinos.
Abstract
The specifications of a newly developed beta-spectrometer, based on full absorption Si(Li) detector and thin transmission detector, are shown. The spectrometer allows to separate effectively beta-radiation and accompanying X-rays and gamma radiation using registration of coincident events from both detectors. The spectrometer can be used for precision measurements of various beta-spectra, namely for the beta-spectrum shape study of ^{144}Pr, which is considered to be an advantageous anti-neutrino source for sterile neutrino searches. The preliminary results for beta spectra from ^{144}Ce-^{144}Pr decays and reconstructed antineutrino spectra are presented.
- Publication:
-
XXVIII International Conference on Neutrino Physics and Astrophysics
- Pub Date:
- June 2018
- DOI:
- 10.5281/zenodo.1300455
- arXiv:
- arXiv:1711.06985
- Bibcode:
- 2018npa..confE.141D
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment;
- High Energy Physics - Phenomenology;
- Nuclear Experiment
- E-Print: