Direct measurement of fast transients by using boot-strapped waveform averaging
Abstract
An approximation to coherent sampling, also known as boot-strapped waveform averaging, is presented. The method uses digital cavities to determine the condition for coherent sampling. It can be used to increase the effective sampling rate of a repetitive signal and the signal to noise ratio simultaneously. The method is demonstrated by using it to directly measure the fluorescence lifetime from Rhodamine 6G by digitizing the signal from a fast avalanche photodiode. The obtained lifetime of 4.0 ns is in agreement with the known values.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- March 2018
- DOI:
- 10.1063/1.5010824
- arXiv:
- arXiv:1709.08389
- Bibcode:
- 2018RScI...89c5104O
- Keywords:
-
- Physics - Data Analysis;
- Statistics and Probability;
- Physics - Instrumentation and Detectors
- E-Print:
- doi:10.1063/1.5010824