Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis
Abstract
Plastic scintillators detectors (PSDs) have been studied as dosimeters, since they provide a cost-effective alternative to conventional ionization chambers. Measurement and analysis of energy dependency were performed on a Siemens Mammomat tomograph for two different peak kilovoltages: 26 kV and 35 kV. Both PSD displayed good linearity for each energy considered and almost no energy dependence.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- January 2018
- DOI:
- 10.1016/j.nima.2017.10.021
- arXiv:
- arXiv:1707.01305
- Bibcode:
- 2018NIMPA.877..346A
- Keywords:
-
- Scintillator;
- Dosimetry;
- Tomosynthesis;
- Photodetector;
- Physics - Instrumentation and Detectors
- E-Print:
- 5 pages, 3 figures, 1 table