Fast x-ray detector system with simultaneous measurement of timing and energy for a single photon
Abstract
We developed a fast X-ray detector system for nuclear resonant scattering (NRS) experiments. Our system employs silicon avalanche photo-diode (Si-APD) as a fast X-ray sensor. The system is able to acquire both timing and energy of a single X-ray photon simultaneously in a high rate condition, 106 counts per second for one Si-APD. The performance of the system was investigated in SPring-8, a synchrotron radiation facility in Japan. Good time resolution of 120 ps (FWHM) was achieved with a slight tail distribution in the time spectrum by a level of 10-9 at 1 ns apart from the peak. Using this system, we successfully observed the NRS from the 26.27-keV level of mercury-201, which has a half-life of 630(50) ps. We also demonstrated the reduction of background events caused by radioactive decays in a radioactive sample by discriminating photon energy.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- June 2017
- DOI:
- arXiv:
- arXiv:1702.04651
- Bibcode:
- 2017RScI...88f3105M
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 7 pages, 12 figures