Comparative Measurements of ITER Nb3Sn Strands Between Two Laboratories
Abstract
ITER Nb3Sn strand quality verification tests require large quantities of precise measurements. Therefore regular cross-checking between testing laboratories is critically important. In this paper, we present results from a cross-checking test of 140 samples between the National High Magnetic Field Laboratory, USA and the University of Twente, the Netherlands. The tests comprise measurements at 4.2 K on critical current, RRR and hysteresis loss, while at room temperature the chromium layer thickness, Cu/nonCu ratio, filament twist pitch, and diameter were determined. Our results show very good agreement between the two labs. The reasons for small random discrepancies are discussed.
- Publication:
-
IEEE Transactions on Applied Superconductivity
- Pub Date:
- August 2017
- DOI:
- arXiv:
- arXiv:1701.00446
- Bibcode:
- 2017ITAS...2785502L
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Condensed Matter - Superconductivity
- E-Print:
- 5 figures