Towards MRI-Based Autonomous Robotic US Acquisitions: A First Feasibility Study
Abstract
Robotic ultrasound has the potential to assist and guide physicians during interventions. In this work, we present a set of methods and a workflow to enable autonomous MRI-guided ultrasound acquisitions. Our approach uses a structured-light 3D scanner for patient-to-robot and image-to-patient calibration, which in turn is used to plan 3D ultrasound trajectories. These MRI-based trajectories are followed autonomously by the robot and are further refined online using automatic MRI/US registration. Despite the low spatial resolution of structured light scanners, the initial planned acquisition path can be followed with an accuracy of 2.46 +/- 0.96 mm. This leads to a good initialization of the MRI/US registration: the 3D-scan-based alignment for planning and acquisition shows an accuracy (distance between planned ultrasound and MRI) of 4.47 mm, and 0.97 mm after an online-update of the calibration based on a closed loop registration.
- Publication:
-
arXiv e-prints
- Pub Date:
- July 2016
- DOI:
- 10.48550/arXiv.1607.08371
- arXiv:
- arXiv:1607.08371
- Bibcode:
- 2016arXiv160708371H
- Keywords:
-
- Computer Science - Robotics
- E-Print:
- 14 pages, 7 figures, journal